Chapter 9: Installing, Removing, and Replacing Components
Removing and Installing a Power Supply
276 Quantum Scalar i40 and Scalar i80 User’s Guide
c From the operator panel, select Tools > Diagnostics.
d Continue through the next few screens to run the magazine test
on the replaced magazine.
If a problem is found during testing,
you receive a diagnostic ticket.
See
Running the IVT Diagnostic Subtests Individually – Robotics,
Tape Drive, and Magazine Tests on page 327 more information.
Removing and Installing a Power Supply
The Scalar
®
i40 contains one power supply. The Scalar i80 can have up
to two power supplies for redundancy purposes.
An unused power supply slot must be covered by a filler plate.
This document covers:
Replacing a Power Supply in a Single-Power-Supply System (Scalar
i40 and Scalar i80)
Adding or Replacing a Redundant Power Supply on the Scalar i80
Figure 68 Single Power Supply
System
Scalar i40, single
power supply
Scalar i80, single
power supply
installed in top slot
Filler plate covers
empty power supply
slot