A wrong library configuration is detected: the mailbox storage column is
missing; also error due to drives out of order or multiple drive types
installed.
Check to make sure that the mailbox column is installed and the fiducial label is
properly installed and not damaged. The fiducial label is present on both the
single slot mailbox and the multi-slot mailbox. Only one drive type is installed in
the library with no missing drives.
Contact ATAC
43 xx
An unknown library configuration is detected, the barcode scanner is
unable to read the fiducial label located on the storage column or storage
magazine during a Teach operation.
Ensure that the Microcode level in the Scalar 100 supports the Hardware
installed. Refer to Getting Firmware and Serial Number Information on page
6-40.
Check for a dirty, damaged, missing or wrong fiducial label located at the
storage array in the Library.
Contact ATAC.
44 xx
An unknown fiducial label is detected during a Teach operation.
Ensure that the Microcode level in the Scalar 100 supports the Hardware
installed. Refer to Getting Firmware and Serial Number Information on page
6-40.
Check for a dirty, damaged, missing or wrong fiducial label where the picker is
positioned.
Contact ATAC.
46 xx
The barcode scanner detected a cartridge when one was not expected.
This occurs because the magazine’s fiducial label was not scanned by the
library. Therefore, the library does not recognize the magazine as being
installed in the column.
This error can be caused by a dirty barcode scanner (dust or fingerprints).
Clean the barcode scanner window with a lint free cloth and retry the operation.
Check for a dirty, damaged, missing or wrong fiducial label on the magazine
containing the cartridge in question.
Contact ATAC