Chapter 5 Running Diagnostic Programs
Running the Move Location Test
84 ATL M-Series User’s Guide
15 Repeat steps 13 and 14 as necessary until the desired cycle count
number is displayed.
16 Press
Right until the cursor is all the way to the right and the Select
button appears.
17 Press
Select to set the cycle count, or press Cancel to return to the
Diags: General screen without running the Move Medium test.
The test starts and the GUI displays a
Running Diag screen. The test
stops automatically after completing the selected number of cycles.
When the test is complete, the GUI displays a screen indicating the
test status.
To abort the test before it completes the selected number of cycles,
press
Stop.
18 Press
Back.
The GUI displays the cycle count screen.
19 Press
Cancel, then Back to return to the Diags: General screen.
Running the Move Location Test 5
The Move Location test is similar to the Move Medium test, except that no
cartridges are actually moved.
When the
Move Location test is cycled once, the robotic hand is moved
from its current location to the destination location.
When the
Move Location test is cycled more than once, the robotic hand is
repeatedly moved between the destination location and a home location,
and will return to the home location at the end of the selected number of
Note: Setting the cycle count to
0000 causes the test to run
continuously.